Characteristics
1. The steam can't go through to chamber when the door hasn’t been closed, and the pilot off inform You can process dry operate.
2. When the pressure inner over 0.027MPa, the door be locked and can't be open.
3. If you not choose the sterile time and start the machine, buzzer sound.
4. If you mistake open the door in the course of start, the jacket value auto off and open the chamber value.
5. Lack water the pilot on, heating and sterile pilot off, the buzzer sound and auto release steam.
6. Overload and leakage protection, when leakage or over current occur, it auto off the power.
7. Temperature controlled not work and the inner pressure over 0.165MPa, the safety value auto start opens.
8. The surface of the sterilizer is steel coated, the chamber of the sterilizer is made of stainless steel.
Technical Data
Model technical data | WS-280YDA | WS-200YDA | WS-150YDA |
Sterilizing chamber volume | 280L(φ600×1000mm) | 200L(φ515×1000mm) | 150L(φ440×1000mm) |
working pressure | 0.22 MPa | ||
working temperature | 134℃ | ||
Adjustment of temperature | 40℃ - 134℃ | ||
Time of sterilization | 0-60 min | ||
Time of drying | 0-60 min | ||
Heat average | ≤± 2℃ | ||
Power | 12KW/380V 50Hz | 9KW/380V 50Hz | 9KW/380V 50Hz |
Dimension | 1400×770×1780 | 1400×670×1650 | 1400×600×1300 |
Package | 1680×920×2100 | 1560×820×1850 | 1550×750×1850 |
GW/NW | 465/365 KG | 350/260 KG | 320/240 KG |