Features
1. Manual or Automatic Bolus/Purge/KVO function.
2. Vertical & Horizontal mounting clamp.
3. DPS dynamic pressure display.
4. Anti-bolus system.
5. Large LCD display with adjustable brightness.
6. 2000 events stored can be checked/downloaded.
Technical Specifications
PARAMETER | SPECIFICATION |
Applicable syringes | 5ml,10ml, 20ml, 30ml, 50ml, 60ml of any standard |
Accuracy | +/- 2% with correct calibration |
Rate mode | 0.1ml/h-2000.0ml/h (0.1ml/h or 1ml/h step) |
Time Mode | SOL. VOL: 0.1-9999.9ml (0.1ml step) |
Dose Mode | Volume: 0.1-999.9ml (0.1ml step) |
Total Volume Range | 0.1-9999.9ml (0.1ml step) |
Drug library Mode | (Optional) |
Sequential Mode | (Optional) |
Purge Rate | Programmable and adjustable |
KVO Rate | 0.0~5.0ml/h, Programmable and adjustable |
DPS Occlusion Pressure | LOW: 60(±15) kPa, MEDIUM: 90(±15) kPa HIGH: 120(±15) kPa (Adjustable) |
Alarms | Loose, Occlusion, Empty, Near Empty, Rate Abnormal, Low Battery, No Battery, Infusion Complete, No Action, Syringe Plunger Disengagement |
Alarm Volume | Three levels: LOW, MEDIUM or HIGH |
Bolus Rate | 0.1-2000ml/h, Programmable and adjustable, Automatic/Manual bolus available |
Bolus Volume | Range: 1.0~50.0ml, (0.1ml step) |
Interface | RS-232 |
Brightness | 1~10 levels can be selected |
History Records | 2000 events stored can be checked and downloaded |
Power Source | AC power supply: AC 100 to 240V (±10%), 50/60Hz; DC Power supply: 12~15V, power>15W; Built-in Lithium Battery: Rechargeable, 11.1V, 2000mAh 12 hours of continuous operation |
Protection Class | Class I, Type CF, IPX4 |
Dimension/N.W. | 240mm*150mm*117mm, 2.5kg |
Packing Size/G.W. | 200mm*260mm*160mm, 3.5kg |